74
BG 5195 BEN / A (2011-05) TPG261.oi
Transmit: SCT [,x] <CR>[<LF>]
│
└─ Display channel, x =
0 –> Gauge 1
1 –> Gauge 2
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x <CR><LF>
│
└─ Display channel
Transmit: ERR <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: xxxx <CR><LF>
└─ xxxx =
0000 –> No error
1000 –> Error Controller error
(See display
on front panel)
0100 –> NO HWR No hardware
0010 –> PAR Inadmissible
parameter
0001 –> SYN Syntax error
The ERROR word is cancelled when read out.
If the error persists, it is immediately set again.
Measurement
channel change
Error status