EasyManua.ls Logo

Pfeiffer TPG 261 - Page 75

Pfeiffer TPG 261
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BG 5195 BEN / A (2011-05) TPG261.oi 75
Transmit: RES [,x] <CR>[<LF>]
└─ x = 1 –> Cancels currently active
error and returns to
measurement mode
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: [x]x,[x]x,... <CR><LF>
└─ List of all present error messages,
xx =
0 –> No error
1 –> Watchdog has responded
2 –> Task fail error
3 –> EPROM error
4 –> RAM error
5 –> EEPROM error
6 –> DISPLAY error
7 –> A/D converter error
9 –> Gauge 1 error (e.g. filament
rupture, no supply)
10 –> Gauge 1 identification error
11 –> Gauge 2 error (e.g. filament
rupture, no supply)
12 –> Gauge 2 identification error
Reset