EasyManua.ls Logo

PFlow Industries D116 - Page 26

PFlow Industries D116
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision: 3.0.3
Holding with Poor Sig
Select "Yes"
to hold last good flow signal displayed if
the Flowmeter experiences a poor signal condition.
This function will allow
to calculate flow totalizer
data without interruption. Select "NO",
instead.
Empty Pipe Setup
This parameter is used to overcome the possible
problems that usually show up when the pipe being
measured is empty. Since signals can be transmitted
through the pipe wall, the flow meter
may still read a
flow while measuring an empty pipe. To prevent this
condition from happening, you can specify a value.
When the signal quality falls below this value, the
measurement stops automatically. If the
flow meter
already able to stop measuring when the pipe is empty,
a value in the range of 30 to 40 should als
o be entered
in this window to ensure no measurement when the
pipe is empty.
Measurement Units Options
Select the measurement unit as follows:
0. Metric
1. English
Flow Rate Units Options
The following flow rate units are available:
0. m3 Cubic Meters
1. 1 Liters
2. gal USA Gallons
3. ig Imperial Gallons
4. mg Million Gallons
5. cf Cubic Feet
6. bal USA Barrels
7. ib Imperial Barrels
8. ob Oil Barrels
The following time units are available:
/ Day / Hour
/ Min / Sec
Factory default is Cubic Meters/hour
Menu 2 8
Menu 2 9
Menu
3
0
Menu
3
1
D116
Series Ultrasonic Flowmeter
to hold last good flow signal displayed if
the Flowmeter experiences a poor signal condition.
to calculate flow totalizer
instead.
This parameter is used to overcome the possible
problems that usually show up when the pipe being
measured is empty. Since signals can be transmitted
may still read a
flow while measuring an empty pipe. To prevent this
condition from happening, you can specify a value.
When the signal quality falls below this value, the
flow meter
is
already able to stop measuring when the pipe is empty,
o be entered
in this window to ensure no measurement when the
Holding with
PoorSig
NO
Empty Pipe Setup [29
0
Measurement Units In
0. Metric
Flow Rate Units [31
m3/h
Series Ultrasonic Flowmeter
Page 26 of 53
PoorSig
Empty Pipe Setup [29
Measurement Units In
Flow Rate Units [31

Table of Contents