Revision: 3.0.3
Set Static State Zero
When fluid is in the static state, the displayed value is
zcalled "Zero Point". When "Zero Point
in the Flowmeter, the
difference is going to be added
into the actual flow values and measurement
Differences will occur in the Flowmeter.
Setting
zero must be carried out after the transducers
are installed and the flow inside the pipe is in the
absolute static state. Thus, the "Zero Point"
Reset Zero
Select "YES"; reset "Zero Point"
user.
Manual Zero Point
This method is
not commonly used. It is only suitable
for experienced operators to set zero under conditions
when it is not preferable to use other methods. Enter
the value manually to add to the measured value to
obtain the actual value. For example:
Actual measured value=250 m3/H
Actual measured value=10 m3/H
Flowmeter Display=240 m3/H
Normally, set the value as "0".
Scale Factor
The scale factor named as instrument
to modify the measurement results.
a numerical value according to the actual
results.
Network IDN
Input system identifying code, these numbers can
be selected from 1~247 except
TER), 10 (0AH Newline), 42 (2AH﹡
6H&) are reserved.
System IDN is used to identify the flowmeter to
a network.
Menu 4 2
Menu 4
3
Menu 4 4
Menu 4 5
Menu 4 6
D116
Series Ultrasonic Flowmeter
When fluid is in the static state, the displayed value is
difference is going to be added
into the actual flow values and measurement
.
zero must be carried out after the transducers
are installed and the flow inside the pipe is in the
not commonly used. It is only suitable
for experienced operators to set zero under conditions
when it is not preferable to use other methods. Enter
the value manually to add to the measured value to
is used
Input system identifying code, these numbers can
System IDN is used to identify the flowmeter to
NO
0
1
88
Series Ultrasonic Flowmeter
Page 29 of 53