CO2 Response Time for Infant/Neonatal Accessories
Specifications
9-18 Efficia CM Series Rack and Modules Instructions for Use
CO
2
Response Time for Infant/Neonatal Accessories
IBP Specifications
Part Number Description
Rise Time
(typical)
Delay
Time
(typical)
Total
System
Time
(typical)
989803159581 VitaLine H Set intubated circuit 240 ms 5.3 s 6.5 s
989803160261 FilterLine™ H Set Long intubated
circuit
240 ms 3.0 s 7.2 s
989803178001 CapnoLine™ H O2 non-intubated
dual-purpose circuit
Up to 310 ms Up to 5.9 s 7.2 s
989803178011 CapnoLine™ H Long non-
intubated single-purpose circuit
Up to 320 ms Up to 5.9 s 7.2 s
989803178021 Nasal FilterLine™ non-intubated
ventilation single-purpose circuit
Up to 250 ms
(standard length)
Up to 4.0 s 5.3 s
M1923A FilterLine™ H Set intubated circuit 180 ms 2.7 s 3.9 s
M4691A CapnoLine™ H non-intubated
single-purpose circuit
Up to 250 ms Up to 4.0 s 5.3 s
Parameter Specification
Measurement Range -40 mmHg – 360 mmHg
Input Sensitivity 5m V/V/mmHg
Zero Static Offsets Up to ± 200 mmHg with ± 1 mmHg accuracy
Transducer Load Impedance Memscap Load impedance: 200 to 2000 Ohms
Memscap Output impedance:
< 3000 Ohms
Transcap Load impedance: 300 to 700 Ohms
Transcap Output impedance: 300 to 1,000 Ohms
Overall Accuracy (including transducer) ± 4% of reading or ± 4 mmHg, whichever is greater
Volume displacement of CPJ840J6 0.2 mm
3
/100 mmHg
Warmup time of equipment and transducer < 15 seconds.