Care and Cleaning
Efficia CM Series Rack and Modules Instructions for Use 7-1
7
Care and Cleaning
To clean or disinfect your Efficia rack and measurement modules, use only the approved
cleaning agents listed in this chapter.
For information on how to clean or disinfect accessories, see the instructions for use
provided with the accessory.
General Guidelines
Keep the rack, modules, cables, and accessories free of dust and dirt. Periodically inspect
vents for dust and clean the vents if necessary. After you clean and disinfect the
modules, check them carefully and do not use it if you see signs of deterioration or
damage.
If you need to return any equipment to Philips, clean and disinfect it first.
Cautions • It is the responsibility of the hospital to determine the appropriate cleaning protocol
for each care unit.
• Do not use unapproved cleaning or disinfecting agents. Even small quantities of some
cleaning agents will damage the monitor.
• Do not use abrasive cleaners or strong solvents such as acetone or acetone-based
compounds. The warranty does not cover damage caused by using unapproved
substances.
• Follow these general precautions:
• Always dilute cleaning agents according to the manufacturer’s instructions or
use the lowest possible concentration.
• Do not allow liquid to enter the monitor case or modules. Do not immerse any
part of the equipment in liquid.
• Do not pour liquid onto the system. If you spill liquid on the exterior of the monitor,
use a clean cloth to dry the monitor. If you believe the liquid may have gotten inside
the monitor, contact your technical support team, who can verify the performance and
safety of the unit.
• Do not use abrasive material (such as steel wool or silver polish).
• Do not autoclave, steam sterilize, or ultrasonically clean the monitor or cables.
• Do not use bleach on the electrical contacts or connectors.