Sensor Application
Monitoring Bispectral Index (BIS™)
5-8 Efficia CM Series Rack and Modules Instructions for Use
Sensor Application
Changing BIS Settings
Changing the Sweep Speed
Step
1
If you are using the Extended Sensor, using a ballpoint pen, record the date
and time of application on the sensor.
2
Wipe skin with alcohol, and dry.
3
Apply sensor on forehead as follows:
Circle #1: Center of forehead, approximately 2 inches (5 cm) above nose.
Circle #4: Directly above and parallel to eyebrow.
Circle #3: On the temple area between corner of eye and hairline.
4
Press edges of sensor to ensure adhesion.
Circle all four elements/circles to seal in gel.
5
Press each circle for five seconds to ensure
proper contact.
Use fingertip
Press circles one at a time
Press firmly
6
Insert sensor tab into patient interface cable until fully engaged.
7
If you are using the Extended Sensor, replace the sensor within 24 hours.
When disconnecting sensor, press release button on patient interface cable.
Step
1
Select any waveform.
The waveform menu appears.
2
Select Sweep Speed from the Select Sweep Speed menu.
3
Select a sweep speed option. Available options are:
6.25 mm/s
12.5 mm/s
25.0 mm/s
50.0 mm/s