Monitoring Anesthetic and Respiratory Gases
Efficia CM Series Rack and Modules Instructions for Use 6-11
Configuring Waveform Settings
Configuring Apnea Alarms
The Apnea alarm time defines the length of time between breaths before generating the
Apnea alarm. For example, if the Apnea Alarm time is set to 20 seconds, and the monitor
does not detect a breath for more than 20 seconds, the monitor sounds an Apnea alarm.
Apnea options are:
• 20 seconds
• 25 seconds
• 30 seconds
• 35 seconds
• 40 seconds
• 45 seconds
• 50 seconds
• 55 seconds
• 60 seconds
Note: The apnea algorithm starts after 4 breaths are taken by the patient.
Changing measurement units
Depending on what type of measurement you select, will determine which Units you
have available to select. The following table show which units are available for each
measurement type.
Step
1
Open the waveform menu.
2
Select the Apnea Time field.
3
Select the Apnea time.
4
Close the window.
Measurement Unit
O2 %
CO2 mmHg,%, kPa
N2O %, MAC
AGT %, MAC
AGT2 %, MAC