Using the Efficia Rack and Modules Safely
Basic Operation
2-4 Efficia CM Series Rack and Modules Instructions for Use
Using the Efficia Rack and Modules Safely
Modules are inserted into the rack with the front of the module facing out. To insert a
module, gently push the module into the rack slot until you hear the module click into
place.
Module Conflicts
When you connect a module to the rack, if the module conflicts with any of the active
measurements (label conflict, duplicate measurements, etc), the Module Configuration
menu will open automatically. If there is already a window open, the menu will be closed
and the module configuration menu opened. Exceptions include:
• If the Admin/Edit/Discharge Patient Menu is opened or
• If the Date/Time Menu is opened (only when the Date/Time Menu is opened
automatically after a cold start) or
• If the monitor is in Diagnostics mode or
• If the Software Upgrade Menu is open.
Ingress of Liquids Standard
The Efficia CM Series monitor, rack and plug-in modules meet the Ingress of Liquids
Standard IEC 60529, Level IPX1. Patient contact materials comply with ISO 10993.
The Multi-gas CO
2
Connections and dual IBP patient applied parts for use with
the Efficia CM patient monitors and rack are classified as type CF, which
specifies their degree of protection against electrical shock. The BIS patient
applied parts are classified as type BF. All are rated as defibrillator proof.
See the Efficia CM Series Instructions for Use for the complete list of
certifications.
The Efficia CM Series monitor, rack and plug-in modules are suitable in the presence of
electrosurgery.
The monitor and its accessories must be tested by qualified service personnel at regular
intervals to ensure that performance has not been degraded by aging or environmental
conditions. Periodic performance verification tests can be performed, as described in the
Efficia CM Service Guide.
Rack and module replacements may only be made by qualified service personnel.
Warnings Risk of injury or health hazard:
• Explosion Hazard. Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide. Oxygen concentrations must be
<25% and partial pressure <27.5 kPa when no other oxidants are present.
• To prevent or minimize burns caused by a defective neutral electrode connection, use
caution when handling electrodes, transducers, and cables in the presence of high-
frequency surgical equipment. Verify that the equipment is working correctly before
each use.
• Do not open the patient monitor or attempt to change the battery. Incorrect battery
replacement may cause overheating, fire or explosion. If you suspect a problem with
parts within the monitor, contact your technical support team or Philips
representative.