Profile Settings for Rack Modules
Specifications
9-24 Efficia CM Series Rack and Modules Instructions for Use
AGT
2
Scale 0-2.0% 0-2.0% 0-2.0%
General Audible Alarms Yes Yes Yes
Anesthetics’ Audible Alarms Yes Yes Yes
Gas On/Standby On On On
BIS EEG Sweep Speed 6.25 mm/s 6.25 mm/s 6.25 mm/s
BIS Scale -50 to 50 V -50 to 50 V -50 to 50 V
BIS Alarm Audible (Adult) Yes Yes Yes
BIS Alarm Audible (Pediatric) Yes Yes Yes
BIS Alarm Audible (Neonatal) Yes Yes Yes
BIS High Pass Filter 2 Hz 2 Hz 2 Hz
BIS Low Pass Filter 70 Hz 70 Hz 70 Hz
BIS Notch Filter 50 & 60 Hz 50 & 60 Hz 50 & 60 Hz
BIS Bispectra Smoothing Rate 30 seconds 30 seconds 30 seconds
BIS Impedance Check Yes Yes Yes
BIS Pane displayed parameters
BIS, SR, SEF,
SQI, TP
BIS, SR, SEF,
SQI, TP
BIS, SR, SQI
BIS L Pane displayed parameters
*Visible only with a bilateral sensor
BIS L, SR L, SEF
L, SQI L, TP L
BIS L, SR L, SEF
L, SQI L, TP L
BIS L, SR L, SQI
L
BIS R Pane displayed parameters
*Visible only with a bilateral sensor
BIS R, SR R, SEF
R, SQI R, TP R
BIS R, SR R, SEF
R, SQI R, TP R
BIS R, SR R, SQI
R
1. Available on Gas modules without AGT identification.
2. AGT Sweep Speed and Scale apply to all anesthetic agents, independently of the anesthetic agent detected.
Parameter GW ICU OR