Monitoring Anesthetic and Respiratory Gases
Efficia CM Series Rack and Modules Instructions for Use 6-13
Configuring Waveform Settings
about the +/- 75% levels from the center. The actual scale is displayed with each line and
drawn in white. The scale is configurable in the related waveform configuration dialog.
To view and/or change the default alarm limits for the Gas measurement module, select
the numeric pane on the right side of the screen that corresponds with the waveform you
are measuring to display the Gas Menu. The General tab of the Gas Menu includes the
following selections and default limits:
There are between three and five parameter selection boxes. Select one parameter for
each box. These parameters will be displayed on the numeric pane on the main screen of
the patient monitor.
*If using the dual agent module, there will be inAGT2 and etAGT2 also.
Measurement Options
awRR (rpm) Upper 30
Lower 8
etCO2 (mmHg) 60
25
imCO2 (mmHg) 4
etO2 (%) 100
10
inO2 (%) 100
18
Apnea Time Default is 20 seconds
CO2 Unit mmHg,%, kPa
Gas On/Standby On
Off
See “Gas Standby Mode”
on page 6-15
O2 Compensation
(available only for modules
without O
2
)
41—60%
Parameter Selections
awRR
imCO2 (inspired minimum CO2)
etCO2
(end tidal carbon dioxide)
inN2O
etN2O (end tidal nitrous oxide concentration)
inO2
etO2
(end tidal oxygen concentration)
inAGT (inspired agent)*
etAGT (expired agent)*
MACt