Chapter 8
Alarms and messages
1047358 Rev N Respironics V60 Ventilator User Manual 8-15
Table 8-4: Vent Inoperative alarm messages: summary and troubleshooting
Message Description Corrective Action
Priority
type (ID)
Manually
resettable
Autore-
settable
Silence-
able
Vent Inoperative
1000
3.3 V Supply
Failed
Technical failure. The venti-
lator is in the ventilator inop-
erative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (2) Yes No No
Vent Inoperative
1001
12 V Supply
Failed
Technical failure. The venti-
lator is in the ventilator inop-
erative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (3) Yes No No
Vent Inoperative
1002
Blower Tempera-
ture Too High
Technical failure. The venti-
lator is in the ventilator inop-
erative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (4) Yes No No
Vent Inoperative
1003
Internal Tempera-
ture High
Technical failure of the CPU
PCBA. The ventilator is in the
ventilator inoperative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (5) Yes No No
Vent Inoperative
1004
Internal Tempera-
ture High
Technical failure of the DAQ
PCBA. The ventilator is in the
ventilator inoperative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (6) Yes No No
Vent Inoperative
1005
Internal Tempera-
ture High
Technical failure of the motor
PCBA. The ventilator is in the
ventilator inoperative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (7) Yes No No
Vent Inoperative
1006
Data Acquisition
PCBA ADC Failed
Technical failure. The venti-
lator is in the ventilator inop-
erative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (8) Yes No No
Vent Inoperative
1007
Machine and
Proximal Pres-
sure Sensors
Failed
Technical failure. The venti-
lator is in the ventilator inop-
erative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High (9) Yes No No
Vent Inoperative
1008
Machine and
Proximal Pres-
sure Sensors
Failed
Technical failure. The venti-
lator is in the ventilator inop-
erative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(10)
Yes No No