Planmeca Proline XC Panoramic X-ray 49
TROUBLESHOOTING
User's manual
Er 70 Processor communication error. Turn off the unit first for 30 seconds, try again without
the patient. If the error is still present, please contact your service technician.
Er 71 Generator processor program memory failure. Turn off the unit first for 30 seconds, try
again without the patient. If the error is still present, please contact your service
technician.
Er 72 Keyboard processor program memory failure. Turn off the unit first for 30 seconds, try
again without the patient. If the error is still present, please contact your service
technician.
Er 73 Keyboard processor program error. Turn off the unit first for 30 seconds. If the error is
still present, please contact your service technician.
Er 74 Keyboard processor operating improperly. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 75 Keyboard processor operating improperly. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 80 Keyboard processor EEPROM failure. Turn off the unit first for 30 seconds. If the error
is still present, please contact your service technician.
Er 81 Generator processor EEPROM failure. Turn off the unit first for 30 seconds. If the error
is still present, please contact your service technician.
Er 82 Keyboard processor configuration error. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 83 Generator processor configuration error. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 84 Tube power generator time out. Turn off the unit first for 30 seconds, try again without
the patient and the film cassette. If the error is still present, please contact your service
technician.
Er 90 Keyboard processor EEPROM not activated. Turn off the unit first for 30 seconds. If
the error is still present, please contact your service technician.
Er 91 Keyboard processor stack overflow. Turn off the unit first for 30 seconds. If the error is
still present, please contact your service technician.
Er 99 Error code generation error. Turn off the unit first for 30 seconds. If the error is still
present, please contact your service technician.
ERROR CODE ERROR MESSAGE EXPLANATION