CLEANING
50 Planmeca Proline XC Panoramic X-ray
User's manual
14 CLEANING
NOTE When disinfecting the unit surfaces, always
disconnect the unit from mains.
The bite piece and the patient support handles can be
cleaned with alcohol-based solutions. The bite piece
and patient support handles must be cleaned after
every patient.
Other unit surfaces can be cleaned with a soft cloth
damped in a mild cleaning solution. It is recommended
to clean the surfaces once a day.
15 SERVICE
To guarantee user and patient safety and to ensure
image quality the unit must be checked and
recalibrated by a qualified PLANMECA service
technician once a year or after every 10,000
exposures if this is sooner. Please refer to the
Planmeca Proline XC Panoramic X-ray technical
manual for complete servicing information.