WIRING DIAGRAM FOR EXISTING PERMANENTLY
INSTALLED POWER CABLE
WARNING
Ensure that the power cable is not connected to live voltage.
The X-ray unit main switch does not cut off the voltage in the mains terminal block.
NOTE
The quality and cross-sectional area of the existing power cable must fulfill the
requirements of the local standards (must be e.g. IEC approved (CE marked),
UL / CSA approved).
L
N
ProOne Permanent Installation
06.06.2016 LY
ProOne
Input
Module
N
LN
Existing power cable