Lamination of image plate becoming detached at edge
Possible cause Remedy
Incorrect retainer system used Only use original image plates and film retainer
systems.
Image plate handled incorrectly Use the image plate correctly.
Observe the operating instructions for the image
plates and film retainer systems.
X-ray image shows pre-erasure at one end
Possible cause Remedy
After the light protection cover has been torn off
and before pushing into the input unit, the image
plate is pushed out of the light protection cover
Do not push out the image plate until the torn-off
light protection cover has been placed on the
input unit.