INCISOR EXPOSURE
24 Planmeca ProX
User’s Manual
11 INCISOR EXPOSURE
11.1 Selecting the exposure parameters
The preprogrammed exposure values are shown in
section 15 “EXPOSURE VALUES” on page 34.
Check that you are in the desired mode: in the film-based
imaging mode, in the digital imaging mode or in the
phosphoric mode.
The imaging mode can be changed by pressing the MODE
key for 2 seconds.
Select the adult or child mode. The indicator light of the
selected projection will come on.
Select the incisor exposure region with the
preprogrammed setting keys. Press the incisor key once to
select the projection of the maxilla, and press the key twice
to select the projection of the mandible. The indicator light
of the selected projection will come on.
The preprogrammed time, kV and mA values appear on
the respective displays.
The preprogrammed time, kV and mA values can be
temporarily changed with the parameter adjustment keys.
This will not affect the preprogrammed values.
Select the parameter to be adjusted with the SELECT key.
Film-based
imaging mode
Phosphoric
mode
Digital imaging
mode
Child
Adult
Incisors
7
0.08

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV