BITE-WING EXPOSURE
32 Planmeca ProX
User’s Manual
When the parameter value is flashing on the kV display,
the anode voltage can be changed with the parameter
adjusting keys.
When the parameter value is flashing on the mA display,
the anode current can be changed with the parameter
adjusting keys.
When the parameter value on the kV or mA display is not
flashing, the exposure time value can be changed with the
parameter adjusting keys.
NOTE After adjusting the kV or mA value the unit will return
automatically to the time adjustment mode after 5
seconds time.
14.2 Patient positioning
Ask the patient to sit down. Place a protective lead apron
over the patient’s chest.
In the bite-wing exposures the patient closes the teeth
during the exposure on the film’s tab or on the film/sensor
holder.
14.2.1 Positioning the cone
The angle of the cone is indicated on the scale located on
the vertical joint of the tube head.
The optional long cone can be attached into the short
cone. Refer to chapter 7.2 “Selecting the cone” on page 9.
Select the cone angle from the table below.
Position the cone according to the figure below.
TEETH
ANGLE OF
INCLINATION
Bite-wing exposure
Prox_cone_pos.eps
Scale for the cone angle
Bite-wing

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV