Planmeca ProX 35
EXPOSURE VALUES
User’s Manual
NOTE The values in the following tables correspond to the
density value 0.
When using the 30 cm long cone program the values
according to the table given in section 17 “EXPOSURE
VALUE TABLES” on page 40 or select three steps darker
density (longer exposure time).
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV mA time kV mA time kV mA time
Adult Maxilla 60 7 0.100 63 6 0.1 63 6 0.125
Mandible 60 7 0.08 63 6 0.08 63 6 0.100
Child Maxilla 60 7 0.064 60 7 0.08 60 7 0.100
Mandible 60 7 0.05 60 7 0.064 60 7 0.080
OCCLUSAL
EXPOSURE
ENDODONTIC BITE-WING
kV mA time kV mA time kV mA time
Adult Maxilla 70 6 0.100
60 7 0.100 60 7 0.100
Mandible 70 6 0.100
Child Maxilla 66 6 0.080
60 7 0.080 60 7 0.080
Mandible 66 6 0.080
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV mA time kV mA time kV mA time
Adult Maxilla 60 7 0.200 63 6 0.200 63 6 0.250
Mandible 60 7 0.160 63 6 0.160 63 6 0.200
Child Maxilla 60 7 0.125 60 7 0.160 60 7 0.200
Mandible 60 7 0.100 60 7 0.125 60 7 0.160
OCCLUSAL
EXPOSURE
ENDODONTIC BITE-WING
kV mA time kV mA time kV mA time
Adult Maxilla 70 6 0.200
60 7 0.200 60 7 0.200
Mandible 70 6 0.200
Child Maxilla 66 6 0.125
60 7 0.160 60 7 0.160
Mandible 66 6 0.125

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV

Related product manuals