40 P/N 2018 - 10 B Rev A 6 – Release
EMC Table
Emission Guidance
The QuickSee is intended for use in the electromagnetic environment specified as follows. The opera-
tor/user of the device should ensure that the device is used in the appropriate environment.
RF Emission
Cispr 11
Group 1
The QuickSee emits electromagnetic energy
in order to function. Nearby electronic equip-
ment may be affected.
RF Emission
Cispr 11
Class A compliant
The QuickSee is suitable for use in all estab-
lishments other than domestic establishments
and those directly connected to the public
low-voltage power supplies building used for
domestic purposes.
Emission
Test
Test Level
IEC 60601-1-2
Conformity
Level
Electromagnetic
Environment Guide
Conducted
Emissions
CISPR 11:2015
+A1:2016
Frequency
(MHz)
Quasi-Peak
Limit dB (μV)
Avg.
Limit dB (μV)
Measured
Quasi-Peak
Limit dB (μV)
Measured
Avg.
Limit dB (μV)
Ambient temperature: 21 °C
Relative humidity: 41 %
0.15 to 0.50
79 66 N/A N/A
0.50 to 30
73 60 N/A N/A
Radiated
Emissions
CISPR 11:2015
+A1:2016
30 to 230
52 to 45 N/A N /A N /A
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range
230 to 1000
52 N/A N /A N /A
Harmonic
Emissions
IEC 61000-3-2:
2014
Nominal Allowed
Deviation
Deviation According applied Standard,
Clause 1, for systems with
nominal voltage less than
220 V (phase-neutral), the
limits have not yet been
considered. Therefore, test
has been performed only at
230 V / 50 Hz.
Supply
Voltage
(V)
230.00 4.60 1.34
Supply
Frequency
(Hz)
50.00 0.25 0.01
Voltage
Fluctuation/
Flicker Emissions
IEC 61000-3-3:
2013
Plt dmax
N/A
Ambient temperature: 22 °C
Relative humidity: 40 %
Limit
0.650 6