PRODIGIT 3260A Series Operation Manual 41
4-5 GPIB/RS-232C command description
4.5.1 Setting Command
CURR
Purpose
The load current setting in Constant Current mode.
Command syntax
CURR{LOW|HIGH|A|B}{SP}{NR2}{NL}
Description
This command is used to set the load current LEVEL A/B of 3260A series Electronic
Load.
Note:
1. The load current data must include decimal point, otherwise this command is
unable to execute. The effective load current level can be programmed is the sixth
digit after the decimal point.
2. The HIGH load level load current MUST be higher than LOW level load current for
proper dynamic waveform definition, the load current difference between
programmed HIGH and LOW level is 10 times of resolution at each range, the
3260A series Electronic Load will adjust and limit of the programmed values.
3. If the programming loads current level over the maximum specification at
programmed range of 3260A series load, the full scale current will be sent to the
load module.
4. Engineering unit for load current is Amps.
For Example
CURRA 1.8 ; Set level A load current to 1.8 A.
CURRB 15.123456 ; Set level B load current to 15.123456 A.
RES
Purpose
The load resistance setting in Constant Resistance mode.
Command syntax
RES{LOW|HIGH|A|B}{SP}{NR2}{NL}
Description:
This command is used to set the load resistance for LEVEL A/B of 3260A series
electronic load.
Note
1. The load resistance data must include decimal point, otherwise this command is
unable to execute. The most effective load resistance level can be set is the sixth
digit after the decimal point.
2. If the programming loads resistance level over the maximum specification at
programmed range of 3260A series load, the full scale resistance will be sent to
the load module.
3. The engineering unit for load resistance is Ohms.
For Example
RESA 9.123 ; Set level A load resistance to 9.123 OHM.
RESB 13.456789 ; Set level B load resistance to 13.456789 OHM.