PRODIGIT 3260A Series Operation Manual 5
1-5 Specification
MODEL 3261A 32611A 32612A 32613A 32614A 32615A
LOAD INPUT RATINGS
Power (VA)
Current(Ampere)
Voltage(Volt)
1800 VA
18 Arms
50~300 Vrms
3600 VA
36 Arms
50~300 Vrms
5400 VA
54 Arms
50~300 Vrms
7200 VA
72 Arms
50~300 Vrms
9000 VA
90 Arms
50~300 Vrms
10800 VA
108 Arms
50~300 Vrms
PROTECTION:
Over Power Protection
Over Current Protection
Over Voltage Protection
Over Temp. Protection
≒ 1890 VA
≒ 18.9 A
≒ 315 V
85℃
≒ 3780 VA
≒ 37.8 A
≒ 315 V
85℃
≒ 5670 VA
≒ 56.7 A
≒ 315 V
85℃
≒ 7560 VA
≒ 75.6 A
≒ 315 V
85℃
≒ 9450 VA
≒ 94.5 A
≒ 315 V
85℃
≒ 11340 VA
≒ 113.4 A
≒ 315 V
85℃
Range
0~9/9~18
A
0~18/18~36
A
0~27/27~54
A
0~36/36~72
A
0~45/45~90
A
0~54/54~108
A
Resolution
2.25/4.5 mA
4.5/9 mA 6.75/13.5 mA
9/18 mA
11.25/21.5 mA
13.5/27 mA
Accuracy ±( 0.5﹪ of setting + 1﹪ of range );
±0.5﹪ of (setting + range)@ 50/60Hz
0~0.9 A 0~1.8 A 0~2.7 A 0~3.6 A 0~4.5 A 0~5.4 A
CC MODE
&
Linear CC
MODE
Low
Current
Accuracy
± 2 % of (setting + range)
Range II/I
3.333~13.33
2~53.332K
1.667~6.668
~26.668K
1.111~4.444
~17.776K
0.833~3.333
~13.33K
0.666~2.666
~10.666K
0.556~2.224
~8.888K
Resolution
0.019/0.076
mS
0.037/0.148
mS
0.056/0.224
mS
0.075/0.300
mS
0.0937/0.375
mS
0.113/0.452
mS
Accuracy ( 0.5﹪ of setting + 2﹪ of range );
±0.5﹪ of (setting + range)@ 50/60Hz
CR MODE
Under 5 ﹪
F.S.Current
±2﹪ of (setting + range)
Range
√2 ~ 3.5 / 1.5 ~ 1.9 / 3.0 ~ 3.4
CREST
FACTOR
(CCMODE
ONLY)
Resolution
0.5 / 0.1 / 0.1
VOLTAGE Range 300 V 300 V 300 V 300 V 300 V 300 V
READBACK
Resolution
0.1 V 0.1 V 0.1 V 0.1 V 0.1 V 0.1 V
V METER Accuracy
±( 0.5﹪ of reading + 0.2﹪of range )
DVM under 1% don’t display.
CURRENT Range 18 A 36 A 54 A 72 A 90 A 108 A
READBACK
Resolution
0.001 A 0.01 A 0.012 A 0.012 A 0.01 A 0.012 A
A METER Accuracy
±( 0.5﹪ of reading + 2﹪ of range );
±0.5﹪ of (reading + range)@ 50/60Hz
WATT Range 1800 W 3600 W 5400 W 7200 W 9000W 10800 W
READBACK
Resolution
0.1 W 1 W 1.2 W 1.2 W 1W 1.2 W
Accuracy ±( 0.5﹪ of reading + 2﹪ of range );
±0.5﹪ of (reading + range)@ 50/60Hz
W METER
Under 5 ﹪
F.S.Current
±3﹪ of (setting + range)
VA METER Vrms×Arms Correspond To Vrms and Arms
FREQUENCY Range DC , 40~400 Hz(CC Mode) ; DC , 0.1~400 Hz(LIN,CR Mode)
Imonitor (Isolated) 4.5 A/V 9 A/V 13.5 A/V 18 A/V 22.5A/V 27 A/V