PRODIGIT 3260A Series Operation Manual 43
PRES
Purpose
Set the upper or lower 4 1/2 digit multi-function meter to display the programming
load level.
Command syntax
PRES{SP}{0|1|OFF|ON}{NL}
Description
PRES ON set the load to preset on status.
MODE
Purpose
Select the operating mode of Electronic load module.
Command syntax
MODE{SP}{0|1|2|CC|CR|LIN}{NL}
Description
MODE CC set the presently operating mode to Constant Current mode for 3260A
Series Load.
MODE CR set the presently operating mode to Constant Resistance mode.
MODE 1 set the presently operating mode to Constant Resistance mode.
STORE
Purpose
STORE the load level and load status into the non-volatile memory of the 3260A
series electronic LOAD.
Command syntax
STORE{SP}{1|2|3|4|5}{NL}
Description
5 different states with 3260A series electronic load's status and load current into the
non-volatile memory in the 3260A series electronic load.
Note:
The new load level and load status can overcome the load level and status if
STORE new state number, because the memory location is the same for same
STORE state number.
For Example
STORE 1
STORE 5