32
MARK76V3002 Rev 3
Recommended separation distances between portable and mobile RF
communications equipment and the ProSun V3 Model 42V3 Non life supporting ME
Equipment
The ProSun V3 Model 42V3 is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the ProSun V3 Model 42V3 can
help prevent electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the ProSun V3 Model 42V3 as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,7 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1.0 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaectedby
absorptionandreectionfromstructures,objectsandpeople.