EasyManua.ls Logo

Prusa i3 MK3 - C Alibrate​ ​ Z

Prusa i3 MK3
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
At the end of the X/Y calibration, the printer measures the reference height above
each of the 9 bed sensor points and stores the reference heights in nonvolatile
memory. During normal bed leveling, it is expected that the PINDA probe triggers no
more than 1 mm from the reference value, therefore the nozzle is not allowed to
move more than 1 mm below the reference value during the bed calibration.
If you moved the printer, you may need to re-run the Z calibration to sample new
reference Z height values reflecting the twist and bend of the table surface the printer
is sitting on. If that does not help, please verify, that the PINDA probe is aligned with
the sensor points on the print bed during the bed Z calibration. The alignment shall
be ensured by the automatic X/Y calibration routine. If the PINDA probe is no more
aligned during the Z calibration over time, it is possible, that a pulley is slipping or
something on the machine frame got loose.
3) Bed
leveling
failed.
Sensor
triggered
too
high.
Waiting
for
reset.
Similar to case 2). This time the PINDA sensor triggered more than 1 mm above the
reference height. Before doing anything else, make the Z axis level by going all the
way up and try again.
6.3.6 Calibrate Z
Calibrate Z is located in the Calibration menu. It is always done with the steel sheet on.
It should be performed whenever you move the printer to a different location. It saves the
heights of all 9 calibration points in nonvolatile memory. Stored information is used every
time mesh bed leveling is called during a print. When the measured values are vastly
different from the stored values, the print is canceled because it is a good indicator
something is wrong. Calibrate Z is a part of the routine Calibrate XYZ  so there’s no need to
run it after a successful Calibrate XYZ.
It is a good practice to run this procedure every time you travel with the printer or the printer
is shipped  the geometry might change slightly and cause an error.
At the start of the Z calibration procedure the printer homes X and Y axis. After that, Z axis
will begin to move up until both sides touch the printed parts at the top.
Please make sure the Z axis went all the way up and you heard a rattling sound as the Z
stepper motors skip steps. This procedure ensures, that 1) the X axis is perfectly horizontal,
2) the print nozzle is in a known distance from the print bed. In case the Z carriage did not
touch the end stoppers, the printer could not possibly know the height of the print nozzle
above the print bed and it could therefore crash into the print bed during the Z calibration
procedure.
The Z calibration procedure also prompts you to "Please
clean
the
nozzle
for
calibration.
Click
when
done."
20

Table of Contents

Related product manuals