EasyManuals Logo

QuadTech Sentry 30 Plus User Manual

Default Icon
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Operation Page 35 of 85
2.4 Programming a Ground Continuity Test
This test is applicable to the Sentry 10, 20 and 30 Plus instruments. A Ground Continuity (GC)
test is usually done first to verify the ground connection before high voltage is applied in the AC
& DC Hipot tests. With the instrument in ‘stand-by’ status, press [F1] = PROGRAM. The AC
test mode programming screen is displayed. Follow the green arrows (!) on the right side of
this diagram to program the individual GC test parameters.
Continue on Next Page.
Set
Current
UP
DOWN
EXIT
NEXT
RMT ERROFSTLOCKSelect Mode
F1
F4
F3
F2
LOWGCSTEP 1/1
CURR:
HIGH:
DWELL:
0.0 A
1.0
0.3 s
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0.1A
F1
F4
F3
F2
LOWGCSTEP 1/1
CURR:
HIGH:
DWELL:
0.0A
1.0
0.3s
To select Mode = GC
To move highlighted box to CURR.
Set
HIGH
Resistance Limit
To set current to 0.1A
To move highlighted box to HIGH.
OFF
OFF
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0.1 - 5
F1
F4
F3
F2
LOWGCSTEP 1/1
CURR:
HIGH:
DWELL:
0.1A
1.0
0.3s
To set High resistance limit:
0.1 - 5
in 0.1
increments.
OFF
Select
GC
Mode
:
:
:
To move highlighted box to DWELL.
In STAND BY:
Press [F1] =
PROGRAM
To move highlighted box to AC.
MORE..
EXIT
NEXT
RMT ERROFSTLOCKVoltage is 0
F1
F4
F3
F2
FALL
RAMP
ARC
LOWACSTEP:
VOLT:
HIGH:
TIME:
0.000 kV
0.500 mA
3.0 s
OFF
OFF
OFF
OFF
:
:
:
:
1/1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the QuadTech Sentry 30 Plus and is the answer not in the manual?

QuadTech Sentry 30 Plus Specifications

General IconGeneral
BrandQuadTech
ModelSentry 30 Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals