Page 70 of 85 Interface
3.2 G16 International Power Strip
The Sentry Plus Series instruments can be connected to the G16 International Power Strip as
illustrated in Figure 3-6 for safety testing of many European corded products.
*Australia *United Kingdom *Denmark
*North America *Norway *Finland
*Sweden *Germany *Netherlands
*Austria *Switzerland *Italy
1.
Connect the three G-16 ground connectors to the Sentry Plus RTN/LOW terminal.
2.
Note: there is a second RTN/LOW terminal on the rear of the Sentry instrument.
3.
Connect the white banana plug to the Sentry Plus OUTPUT terminal.
4.
Plug DUT’s AC power cord into corresponding module on G16 power strip.
Figure 3-6: G16 International Power Strip Connection to Sentry 30 Plus
G16 International Power Strip
Black
Banana
Plug
to
RTN
Green
Lug
to
RTN
Green/
Yellow
Lug
to
RTN
or RTN
on Rear
Panel
White
Custom
Banana
Plug
to
HV
OUTPUT
F1
RTN/LOW
F2
OUTPUT
DANGERPASS FAIL
F1
F2
F3
F4
START
STOP
Sentry 30 Plus AC/DC/IR Hipot Tester
01
Q
uadTech
!
CAUTION
F4
F3
Max 5kVAC
6kVDC
CAL UPDATE
QuadTech
AUSTRALIA UNITED KINGDOM DENMARK NORTH AMERICA
NORWAY FINLAND
SWEDEN
SWITZERLAND ITALY
GERMANY
AUSTRIA NETHERLANDS
Sentry 30 Plus Hipot Tester

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the QuadTech Sentry 30 Plus and is the answer not in the manual?

QuadTech Sentry 30 Plus Specifications

General IconGeneral
BrandQuadTech
ModelSentry 30 Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals