Page 82 of 85 Service & Calibration
4.3.6 AC Current Calibration (ACA)
Connect the OUTPUT terminal of the Sentry Plus Series instrument to a resistance box or
resistance standard. Connect an AC/DC current meter in series between the resistance load
(box/standard) and the RTN/LOW terminal of the Sentry Plus instrument. Table 4-3 lists the
resistance loads necessary for the current calibration steps.
Table 4-3: Resistance Loads
Mode Step # Voltage Calibration
Point
Resistance
(Load)
ACA CAL 7 1200 V 0.12mA
10MΩ
ACA CAL 8 1200 V 2.4mA full
500kΩ
ACA CAL9 1200 V 2.4mA
500kΩ
ACA CAL 10 1200 V 12mA
100kΩ
DCA CAL 11 1200 V 0.12mA
10MΩ
DCA CAL 12 1200 V 2.4mA full
500kΩ
DCA CAL 13 1200 V 2.4mA
500kΩ
DCA CAL 14 1200 V 4.8mA
250kΩ
Press [F1] = UP to go to Cal Step 7: ‘ACA 3mA Offset (0.12mA)’.
Press [STOP] to return instrument to steady state.
Press [START] to get offset value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 3mA Offset” display (incorporating the new cal value).
Press [F1] = UP to go to Cal Step 8: ‘ACA 3mA Full (2.4mA)’.
Press [STOP] to return instrument to steady state.
Press [START] to get full value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 3mA Full” display (incorporating the new cal value).
Press [F1] = UP to go to Cal Step 9: ‘ACA 15mA Offset (2.4mA)’.
Press [STOP] to return instrument to steady state.
Press [START] to get offset value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 30mA Offset” display (incorporating the new cal value).
Press [F1] = UP to go to Cal Step 10: ‘ACA 15mA Full (12mA)’.
Press [STOP] to return instrument to steady state.
Press [START] to get full value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 30mA Full” display (incorporating the new cal value).