Operation Page 51 of 85
OFFSET Function
There is no offset in the IR or PA modes. Using the pre-programmed example from paragraphs
2.4-2.8 of the GC, AC, DC, IR & PA 5-step test the resultant error and correction is illustrated
herein.
After PASS Screen, press [STOP] and screen reverts to STAND BY display.
Continue on Next Page.
GET
OFFSET
To GET OFFSET values
[F4]
To return to STAND BY status
To enter OFFSET function.
START
MORE..
OFFSET
RMT ERROFSTLOCKSTAND BY
F1
F4
F3
F2
-------------
OUTPUT
1
2.750kV 15.00mA
3
2AC
DC
IR
2.500kV
0.750kV 1.5M
Ω
2.999mA
-------------
-------------
4
5
GC
PA
M01 MEASURE RESULT
0.1A
CHANGE-DUT
1.2
Ω
-------------
-------------
CANCEL
PRESS TEST KEY TO GET OFFSET..
RMT ERROFSTLOCK
F1
F4
F3
F2
1.
2.
PLEASE OPEN THE HV
OUTPUT TERMINAL.
NO OFFSET IN IR MODE.
MORE..
PRESS TEST KEY TO CONTINUE..
RMT ERROFSTLOCK
F1
F4
F3
F2
1.
2.
PAUSE: CHANGE-DUT
UNDER TEST SIGNAL: OFF
M01 5/5 PA
U.T.S = OFF
MORE..
RMT ERR
OFST
LOCK
F1
F4
F3
F2
1.
2.
PAUSE: CHANGE-DUT
UNDER TEST SIGNAL: OFF
M01 5/5 PA
PASS
PASS
OFFSET
VALUES
MORE..
RMT ERROFSTLOCKU.T.S = OFF
F1
F4
F3
F2
PASS
OUTPUT
1
2.753kV 0.07mA
3
2AC
DC
IR
2.499kV
0.750kV 1.5M
Ω
0.002mA PASS
PASS
4
5
GC
PA
M01 MEASURE RESULT
0.1A
CHANGE-DUT
0.1
Ω
SKIPPED
"STAND BY"
Display
with
Memory ID
PROGRAM
PRESET
MORE..
MENU
RMT ERROFSTLOCKSTAND BY
0.1A
0.5s
1.2
Ω
F1
F4
F3
F2
LOWGCSTEP 1/5
To access OFFSET function.
0.3
Ω
M01 :
START