5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results - oximetry, Continued
sO
2
-
macromodes
Bias:
sO
2
(%) ABL835/25/15 ABL830/20/10
ctHb (g/dL) sO
2
(%) S195 FM* S85
15 0 0.00 0.05
−0.02
7 100 0.01 0.22 N/A
15 100 0.01
−0.08
0.00
25 100 0.00
−0.29
N/A
* FM = FLEXMODE (no message) corresponding to C195 mode on the
ABL735/25/15.
Imprecision:
ctHb (g/dL) sO
2
(%) S
0
S
D
S
ABL
S
X
15 0 0.05 0.05 0.25 0.30
7 100 0.10 0.10 0.25 0.30
15 100 0.05 0.10 0.25 0.30
25 100 0.05 0.10 0.30 0.35
sO
2
-
micromodes
Bias:
sO
2
ABL835/25/15 ABL830/20/10
ctHb
(g/dL)
sO
2
(%)
S95 C95 S85 C55 C35 FM* C55 C35
15 0
−0.04 −0.02
-0.02 -0.03 -0.03 N/A
−0.03 −0.03
7 100
−0.10 −0.19
N/A -0.22 -0.10 N/A
−0.22 −0.10
15 100
−0.10 −0.16
0.00 -0.16 -0.10 -0.05
−0.16 −0.10
25 100
−0.10 −0.17
N/A -0.14 -0.09 N/A
−0.14 −0.09
* FM = FLEXMODE (no message) corresponding to C195 mode on the
ABL735/25/15.
Imprecision:
ctHb (g/dL) sO
2
(%) S
0
S
D
S
ABL
S
X
15 0 0.05 0.05 0.25 0.30
7 100 0.10 0.10 0.25 0.30
15 100 0.05 0.10 0.25 0.30
25 100 0.05 0.10 0.30 0.35
Continued on next page
5-32