ABL800 FLEX Reference Manual 5. Performance characteristics
Performance test results - oximetry, Continued
FO
2
Hb -
macromodes
Bias:
FO
2
Hb ABL835/25 ABL830/20
ctHb (g/dL) FO
2
Hb (%) S195 FM* S85
15 0 0.00
0.04 0.02
7 100
0.07
N/A N/A
15 100
0.03
N/A
0.15
25 100
0.05
N/A N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) FO
2
Hb
(%)
S
0
S
D
S
ABL
S
X
15 0 0.05 0.05 0.25 0.30
7 100 0.25 0.20 0.50 0.60
15 100 0.15 0.15 0.45 0.50
25 100 0.10 0.10 0.40 0.45
FO
2
Hb -
micromodes
Bias:
FO
2
Hb (%) ABL835/25
ctHb (g/dL) FO
2
Hb (%) S95 C95 S85 C55 C35
15 0
0.04 0.02 0.02 0.03 0.03
7 100
0.47 0.39
N/A
0.48 0.18
15 100
0.33 0.40 0.15 0.39 0.31
25 100
0.29 0.46
N/A
0.36 0.33
FO
2
Hb ABL830/20
ctHb (g/dL) FO
2
Hb (%) C85 C55 C35
15 0 N/A
0.03 0.03
7 100 N/A
0.48 0.18
15 100
0.16 0.39 0.31
25 100 N/A
0.36 0.33
Continued on next page
5-33