5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results - oximetry, Continued
FO
2
Hb -
micromodes
(continued)
Imprecision:
ctHb (g/dL) FO
2
Hb (%) S
0
S
D
S
ABL
S
X
15 0 0.05 0.05 0.25 0.30
7 100 0.25 0.20 0.50 0.60
15 100 0.15 0.15 0.45 0.50
25 100 0.10 0.10 0.40 0.45
FCOHb -
macromodes
Bias:
FCOHb ABL835/25 ABL830/20
ctHb (g/dL) sO
2
(%) FCOHb
(%)
S195 FM* S85
15 100 0 0.03 0.08 0.12
7 100 20 N/A 0.47 N/A
15 100 20 N/A 0.10 N/A
25 100 20 N/A
0.47
N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) sO
2
(%) FCOHb (%) S
0
S
D
S
ABL
S
X
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
FCOHb -
micromodes
Bias:
FCOHb ABL835/25
ctHb
(g/dL)
sO
2
(%)
FCOHb
(%)
S95 C95 S85 C55 C35
15 100 0 0.10 0.10 0.12 0.08 0.08
7 100 20 N/A N/A N/A N/A N/A
15 100 20 N/A N/A N/A N/A N/A
25 100 20 N/A N/A N/A N/A N/A
Continued on next page