ABL800 FLEX Reference Manual 5. Performance characteristics
Performance test results - oximetry, Continued
FCOHb –
micromodes
(continued)
Bias:
FCOHb ABL830/20
ctHb (g/dL) sO
2
(%) FCOHb (%) FM* C55 C35
15 100 0
0.02
0.08 0.08
7 100 20 N/A N/A N/A
15 100 20 N/A N/A N/A
25 100 20 N/A N/A N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) sO
2
(%) FCOHb (%) S
0
S
D
S
ABL
S
X
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
FMetHb -
macromodes
Bias:
FMetHb ABL835/25 ABL830/20
ctHb (g/dL) sO
2
(%) FMetHb (%) S195 FM* S85
15 100 0 0.01
0.03
0.06
15 100 20 N/A 0.10 N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) sO
2
(%) FMetHb (%) S
0
S
D
S
ABL
S
X
15 100 0 0.10 0.10 0.25 0.30
15 100 20 0.05 0.10 0.35 0.40
Continued on next page