5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results - oximetry, Continued
FMetHb
micromodes
Bias:
FMetHb ABL835/25
ctHb
(g/dL)
sO
2
(%) FMetHb
(%)
S95 C95 S85 C55 C35
15 100 0 0.13 0.14 0.06 0.16 0.14
7 100 20 N/A N/A N/A N/A N/A
15 100 20 N/A N/A N/A N/A N/A
25 100 20 N/A N/A N/A N/A N/A
FMetHb ABL830/20
ctHb (g/dL) sO
2
(%) FMetHb (%) FM* C55 C35
15 100 0 0.13 0.16 0.14
7 100 20 N/A N/A N/A
15 100 20 N/A N/A N/A
25 100 20 N/A N/A N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) sO
2
(%) FMetHb(%) S
0
S
D
S
ABL
S
X
15 100 0 0.10 0.10 0.25 0.30
15 100 20 0.05 0.10 0.35 0.40
FHHb -
macromodes
Bias:
FHHb ABL835/25 ABL830/20
FHHb (%) ctHb (g/dL) S195 FM* S85
0 15 -0.01 0.08
ā0.05
* FM = FLEXMODE (no message)
Imprecision:
FHHb (%)
ctHb (g/dL) S
0
S
D
S
ABL
S
X
0 15 0.05 0.10 0.30 0.35
Continued on next page
5-36