ABL800 FLEX Reference Manual 5. Performance characteristics
Performance test results - oximetry, Continued
FHHb -
micromodes
Bias:
FHHb ABL835/25 ABL830/20
ctHb
(g/dL)
FHHb
(%)
S95 C95 S85 C55 C35 FM* C55 C35
15 0 0.09 N/A N/A 0.15 0.10 N/A N/A 0.10
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) FHHb (%) S
0
S
D
S
ABL
S
X
15 0 0.05 0.10 0.30 0.35
FHbF – adult
blood
Bias (macromodes):
FHbF ABL835 ABL830
FHbF (%) ctHb (g/dL) S195 FM* S85
0 10 3.3 3.3 3.3
0 15 5.5 5.5 5.5
0 20 5.6 5.6 5.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%)
ctHb (g/dL) sO
2
(%) S
0
S
D
S
ABL
S
X
0 10 100 4 4 5 8
0 15 100 2 3 7 8
0 20 100 2 2 10 11
Continued on next page
5-37