5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results - oximetry, Continued
FHbF – adult
blood
(continued)
Bias (micromodes):
FHbF ABL835 ABL830
ctHb
(g/dL)
FHbF
(%)
S95 C95 S85 C55 C35 FM* C55 C35
1
0
0 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3
1
5
0 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5
2
0
0 5.6 5.6 5.6 5.6 5.6 5.6 5.6 5.6
* FM = FLEXMODE (no message)
Imprecision (micromodes):
ctHb (g/dL) FHbF (%) sO
2
(%) S
0
S
D
S
ABL
S
X
10
4 4 5 8
15
2 3 5
7
20
0 100
2 2
10 11
NOTES: a, b.
FHbF – fetal
blood
Bias (macromodes):
FHbF ABL835 ABL830
FHbF (%) ctHb (g/dL) S195 FM*
80 10 5.9 5.9 5.9
80 15 3.3 3.3 3.3
80 20 2.6 2.6 2.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%)
ctHb (g/dL) sO
2
(%) S
0
S
D
S
ABL
S
X
80 10 100 4 5 5 9
80 15 100 3 3 6 8
80 20 100 2 3 6 7
NOTES: a, b.
Continued on next page
5-38

Table of Contents