EasyManuals Logo

RADIOMETER ABL800 FLEX Reference Manual

RADIOMETER ABL800 FLEX
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results - oximetry, Continued
FHbF – adult
blood
(continued)
Bias (micromodes):
FHbF ABL835 ABL830
ctHb
(g/dL)
FHbF
(%)
S95 C95 S85 C55 C35 FM* C55 C35
1
0
0 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3
1
5
0 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5
2
0
0 5.6 5.6 5.6 5.6 5.6 5.6 5.6 5.6
* FM = FLEXMODE (no message)
Imprecision (micromodes):
ctHb (g/dL) FHbF (%) sO
2
(%) S
0
S
D
S
ABL
S
X
10
4 4 5 8
15
2 3 5
7
20
0 100
2 2
10 11
NOTES: a, b.
FHbF – fetal
blood
Bias (macromodes):
FHbF ABL835 ABL830
FHbF (%) ctHb (g/dL) S195 FM*
80 10 5.9 5.9 5.9
80 15 3.3 3.3 3.3
80 20 2.6 2.6 2.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%)
ctHb (g/dL) sO
2
(%) S
0
S
D
S
ABL
S
X
80 10 100 4 5 5 9
80 15 100 3 3 6 8
80 20 100 2 3 6 7
NOTES: a, b.
Continued on next page
5-38

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals