ABL800 FLEX Reference Manual 5. Performance characteristics
Performance test results - bilirubin, Continued
Imprecision
parameters
The following parameters are used to describe performance of the ABL835/30
analyzers for bilirubin measurements.
S
0
: Repeatability. Measurement short time interval variation on the same
sample.
S
D
: Day-to-day variation
S
T
: Patient-to-patient variation
S
I
: ABL-to-ABL instrumental variation
S
ABL
: ABL uncertainty. Variation including S
T
, S
I
and reference uncertainty
S
X
: Reproducibility. Total variation including S
0
, S
D
and S
ABL
The above field test regression statistics S
yx
include variations from S
0
, S
D
and S
T.
Performance
test results for
bilirubin
Macromodes: 195 µL and 85 µL from syringe and capillary:
ctBil
(µmol/L)
ctHb
(g/dL)
sO
2
(%) S
0
S
D
S
T
S
I
S
ABL
S
X
0
Plasma 1.1 1.4 2.2 0.4 2.3 2.9
0
10 100 1.9 3.1 4.0 3.2 5.1 6.3
0
15 100 2.3 2.9 7.4 5.5 9.2 9.9
0
20 100 3.4 2.6 10.9 13.0 17.0 17.5
200
Plasma 1.3 1.7 3.1 4.7 7.4 7.7
200
10 100 2.4 4.4 5.8 6.6 10.1 11.3
200
15 100 2.6 3.7 8.5 9.3 13.6 14.4
200
20 100 4.2 5.0 12.1 15.4 20.4 21.4
400
Plasma 1.7 2.5 4.8 9.3 12.0 12.3
400
10 100 3.5 6.8 9.3 12.0 16.5 18.2
400
15 100 3.4 5.3 11.4 15.9 20.8 21.7
400
20 100 6.0 8.8 15.0 21.0 27.1 29.2
Notes: a, b, c
Continued on next page
5-43

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals