5. Performance characteristics ABL800 FLEX Reference Manual
Interference tests, Continued
Electrolytes
(continued)
Interference on…
Substance Test Conc. cK
+
(4 mmol/L
level)
cNa
+
(150 mmol/L
level)
cCa
2+
(1.25 mmol/L
level)
cCl
(110 mmol/L
level)
Ascorbic
acid
1.0
mmol/L
0
pH 7.2
7.2 0 0 0.01
1
pH 7.6
7.6 0 0
0.01
1
Sulphide will give erroneously high cCl
results.
Metabolites
The following interference results are found for the metabolite electrodes:
Interference on
Substance Test Conc.
(mmol/L)
cGlucose
(4.0 mmol/L level)
cLactate
(1.5 mmol/L level)
Acetoacetic acid 2
<0.1 <0.1
Acetylsalicylic acid 3
<0.1 <0.1
Ascorbic acid 2
<0.1 <0.1
Bilirubin (conjugated) 0.46
<0.1 <0.1
Bilirubin (unconjugated) 0.34
<0.1 <0.1
Chlorpromazine HCl 0.2
<0.1 <0.1
Citrate 50
0.37
0.19
Creatinine 3
<0.1 <0.1
D-glucose 67
<0.1
Dopamine HCl 1.0
<0.1 <0.1
EDTA 3
<0.1 <0.1
Ethanol 79
<0.1 <0.1
Fluoride 50
0.36 <0.1
Galactose 3.3 up to 1.88*
Glucosamine 2 up to 1.06*
Continued on next page
5-48

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals