ABL800 FLEX Reference Manual 5. Performance characteristics
Interference tests, Continued
Metabolites
(continued)
Interference on
Substance Test Conc.
(mmol/L)
cGlucose
(4.0 mmol/L level)
cLactate
(1.5 mmol/L level)
Glycolic acid 1
<0.1
Interference
Heparin 8000 IU/dL
<0.1 <0.1
Ibuprofen 2
<0.1 <0.1
Intralipid
<0.1 <0.1
Lactic acid 12
<0.1
Maltose 5 up to 0.4*
Mannose 1 up to 0.4*
Oxalate 90
0.47
0.14
Paracetamol-4-
acetamidophenol
2
<0.1 <0.1
Pyruvate 2
<0.1 <0.1
Salicylic acid 4
<0.1 <0.1
Thiocyanic acid 24 Interference Interference
Urea 84
<0.1 <0.1
Uric acid 1.5
<0.1 <0.1
Xylose 1 up to 0.34*
* Values determined at cGlu = 0 mmol/L. Interference at cGlu 4.0 mmol/L is
expected to be the same.
cLactate % at :
Hematocrit % 5 mmol/L level 15 mmol/L level
0 0.7 % 0.7 %
45 0.0 % 0.0 %
60
0.5 % 2.0 %
75
2.2 % 5.0 %
Continued on next page