EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 132

RADIOMETER ABL800 FLEX
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Performance characteristics ABL800 FLEX Reference Manual
Interference tests, Continued
The substances against which the oximetry parameters (ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb, FHHb, FHbF) and ctBil were tested for interference are given
in the table below:
Oximetry
parameters
(SAT100 blood reference test sample: ctHb=15 g/dL, sO
2
=100 %, FCOHb=0.7 %,
FMetHb=0.5 %, ctBil=0, pH=7.4. Parameter sensitivity from the influence on the
absorbance spectrum from various substances.)
Change on
Substance Test conc.
ctHb
(g/dL)
sO
2
(%)
FO
2
Hb
(%)
FCOHb
(%)
FMetHb
(%)
FHHb
(%)
FHbF
(%)
ctBil
(
µ
mol/L)
Intralipid 4 Vol %
e)
0.5
0.1
1.3
0.5 0.9
0.1
11 0
4
b)
Intralipid 2 Vol %
f)
0.4
0.1
0.3
0.3 0.1
0.1
11 7
2
b)
HbF
a), c)
20 %
0.02
1.17 0.04 0.73 0.37
1.14
0
14
SHb 10 % 0
1.0
0.9
0.1
0.1
0.9
Not tested
pH 7.1
7.9
0.5
0.6
0.5
0.6
0.2
0.5
0.4
1.0
0.1
0.1
0.5
0.6
19
13
0
5
Cardio Green
c)
5 mg/L
0.16
0.29 1.14 0.07
0.93 0.29 5 20
Evans Blue
c)
5 mg/L
0.04
0.14 0.28
0.20 0.20
0.14
5
5
Betacarotene in
plasma
c)
3.7 µmol/L
0.0
0.02
0.03
0.01 0.04
0.02 0.1
0.2
Patent Blue V
c)
10 mg/L
0.16
0.39 0.86
0.47
0.00
0.38 21
38
Methylene Blue
c)
30 mg/L
0.7 3.4
5.6
3.0 6.2
3.6
37 25
HiCN
c)
0.11
mmol/L
0.26
1.5 3.0 0.5
0.5 1.5 24 47
MCHC
c), d)
newborn range
320 g/L
350 g/L
No interference
12
17
Sedimentation rate
100 arb.
Units
± 0.5
No interference
Not
tested
Notes:
a) If function “Correction for HbF levels less than 20 %” is activated, the change is 0 for all
parameters.
b) Plasma sample.
c) Calculated value from mathematical superposition of measured pure interference spectrum
on measured reference spectrum.
d)
ctBil = 400 µmol/L.
e) Intralipid (20 % solution) at 4 Vol % gives final test level of 0.8 %.
f) Intralipid (20 % solution) at 2 Vol % gives final test level of 0.4 %.
There is no interference from fetal hemoglobin (HbF) when the analyzer applies HbF correction.
There is no interference from bilirubin (conjugated/unconjugated) up to 1000 µmol/L.
Continued on next page
5-50

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals