1. Potentiometric measuring principles ABL800 FLEX Reference Manual
Electrolyte electrodes, Continued
Corrections for cNa
+
(cont): Corrections
(continued)
ABL8XX
FLEX
Mode A
0
A
1
Equation
05 S165 0.995 -3.00 A
S95 1.01 1.80 A, B
C95 1.03 -1.09 A, B
*FM (no message) 0.995 -3.00 A
*FM (message 874) 1.030 -1.00 A, B
*FM = FLEXMODE.
Corrections for cK
+
:
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 0.985 -0.065 A
S95 1.05 -0.13 A, B
C95 1.11 -0.37 A, B
*FM (no message) 0.985 -0.065 A
*FM (message 874) 1.11 -0.37 A, B
05 S165 0.985 -0.065 A
S95 1.05 -0.13 A, B
C95 1.11 -0.37 A, B
*FM (no message) 0.985 -0.065 A
*FM (message 874) 1.11 -0.37 A, B
*FM = FLEXMODE.
Continued on next page
1-30