ABL800 FLEX Reference Manual 1. Potentiometric measuring principles
Electrolyte electrodes, Continued
Corrections for cCa
2+
:
Corrections
(continued)
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 1.004 -0.022 A
S95 1.05 -0.004 A, B
C95 1.08 -0.04 A, B
*FM (no message) 1.004 -0.022 A
*FM (message 874) 1.08 -0.04 A, B
05 S165 1.004 -0.022 A
S95 1.05 -0.004 A, B
C95 1.08 -0.04 A, B
*FM (no message) 1.004 -0.022 A
*FM (message 874) 1.08 -0.04 A, B
*FM = FLEXMODE.
Corrections for cCl
-
:
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 1.225 -30.7 C
S95 1.000 0.0 C, B
C95 1.01 -1.7 C, B
*FM (no message) 1.225 -30.7 C
*FM (message 874) 1.01 -1.7 C, B
05 S165 1.225 -30.7 C
S95 1.000 0.0 C, B
C95 1.01 -1.7 C, B
*FM (no message) 1.225 -30.7 C
*FM (message 874) 1.01 -1.7 C, B
*FM = FLEXMODE.
Continued on next page
1-31