EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 37

RADIOMETER ABL800 FLEX
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Potentiometric measuring principles ABL800 FLEX Reference Manual
Electrolyte electrodes, Continued
Stability criteria
The following stability criterion must be met to obtain a stable electrode response
during calibration:
upd.last)X(Cal,K)upd.iX(Cal,upd.last)X(Cal, ccc ×
This criterion is valid for calibrations using Cal 1 and Cal 2 solutions where:
cX(Cal,upd.last) = Concentration of the electrolyte ion from the last updating
when measuring on calibration solution. (The last
updating is number 30).
cX(Cal,upd.i) = Concentration of the electrolyte ion for a given updating
when measuring on calibration solution. (The relationship
must be fulfilled for at least one of the updating numbers
18 or 19).
K = Constant for the stability criterion.
Electrolyte Ion Cal1 solution Cal2 solution
K
+
0.01 0.01
Na
+
0.01 0.02
Ca
2+
0.02 0.02
Cl
0.022 0.022
The following stability criterion must be met to obtain a stable electrode response
during measurement:
()
)()()upd.lastX(sample,K
)upd.iX(sample,upd.last)X(sample,
RinsecXRinsecXc
cc
+×
where:
cX(sample,upd.last) = Concentration of the electrolyte ion from the median
of the last 5 updatings (for Ca
2+
: 3 last updatings)
when measuring on a sample. The last updating
number is 30 (or 10 for some micromodes).
cX(sample,upd.i) = Concentration of the electrolyte ion for a given
updating when measuring on a sample. (The
relationship must be fulfilled for at least one of the
updating numbers shown below).
K
+
Na
+
Ca
2+
Cl
22 22 26 22
23 23 27 23
In some micromodes, substract 20 from number
above.
Continued on next page
1-32

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals