EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 47

RADIOMETER ABL800 FLEX
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2. Amperometric measuring principles ABL800 FLEX Reference Manual
pO
2
electrode, Continued
Gas/liquid relationship:
Corrections -
blood samples
K
1
is a constant that describes the gas/liquid relationship for the electrode. The
constant is defined as follows:
+++=
66294.3
)O(Sens
712.218370.5
100
1
1K
2
1
p
The pO
2
measured from the sample is then corrected for systematic deviations
from the reference method using the following equation:
Equation A:
(
)
p
dd ee e
O (sample,corr)
O (sample,v1) O (sample,v1)
2
2 2
=
−+ × +× +×
11
2
23 4
2
4 pp
2
where:
pO
2
value of the sample after the first part of correction is as follows:
)398.100()((sample)Ov1)(sample,O
4
2
(sample)O
2122
Bekkpp
pk
3
××+=
×
and:
d
1
= e
0
× pO
2
(sample, v1) + e
1
k
1
= correction constant = 0.02614
k
2
= correction constant = 0.02107
k
3
= correction constant = 0.00281
e
0
, e
1
, e
2
, e
3
, e
4
= correction constants
B
= barometric pressure in kPa
Equation A+:
When an additional correction is needed, equation A is first used together with the
constants for the FLEXMODE (C195 and 165) – no message. Then the obtained
results are put back into equation A as pO
2
(sample) and then treated again, using
the constants for the specific mode.
Or
Equation B:
When an additional correction is needed, equation A is first used together with the
constants for the FLEXMODE (C195 and 165) – no message. Then the obtained
results are put back into equation B as pO
2
(sample) and then treated again, using
the constants for the specific mode.
cX(sample,corr) = A
0
× cX(sample) + A
1
Continued on next page
2-8

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals