ABL800 FLEX Reference Manual 2. Amperometric measuring principles
pO
2
electrode, Continued
Corrections –
blood samples
(continued)
ABL
8XX
FLEX
Mode e
0
e
1
e
2
e
3
e
4
A
0
A
1
Eq.
35/25/ S195
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
15 S95
1.020 -0.200
B
S85
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
C95
0.9965 -0.0254
B
C55
-2.1199 4.7210 -1.0156 -4.2974 1.1035
A+
*FM (no message)
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
*FM (message 874)
0.9965 -0.0254
B
*FM (message 873)
0.9965 -0.0254
B
*FM (message 872)
-2.20159 5.70807 -0.41342 -5.42718 1.19023
A+
*FM (message 871)
-2.1199 4.7210 -1.0156 -4.2974 1.1035
A+
30/20/ S85
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
10 C55
-2.1199 4.7210 -1.0156 -4.2974 1.1035
A+
*FM (no message)
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
*FM (message 872)
-2.19314 5.81012 -0.96320 -5.46921 1.18037
A+
*FM (message 871)
-2.1199 4.7210 -1.0156 -4.2974 1.1035
A+
05 S165
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
S95
1.020 -0.200
B
S85
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
C95
0.9965 -0.0254
B
C55
-2.1199 4.7210 -1.0156 -4.2974 1.1035
A+
*FM (no message)
-2.30300 5.96942 0.83281 -6.07310 1.30565
A
*FM (message 874)
0.9965 -0.0254
B
*FM (message 873)
0.9965 -0.0254
B
*FM (message 872)
-2.20159 5.70807 -0.41342 -5.42718 1.19023
A+
*FM (message 871)
-2.1199 4.7210 -1.0156 -4.2974 1.1035
A+
00
*FM = FLEXMODE.
Continued on next page
2-9

Table of Contents