- 48 -
1- mass of a single part
2 - mass of all counted parts
3 - the pictogram of enabled Automatic Accuracy Correction function
The display indicates mass of counted parts, that are loaded on balance’s weighing pan
(i.e. 10 parts). If the added amount of parts is below the currently counted one, then the
software automatically corrects mass of a single part. In this case it is APW = 5.2282
corrected to 5.1837. from now on the following parts are counted according to the new
mass of a single part.
This means enables counting mass of a single part from a standard quantity.
The software comprises four conditions for operation of
Automatic Accuracy Correction function
1. After adding the number of parts placed on balance’s weighing pan must be
greater than before adding
2. After adding the number of parts loaded on balance’s weighing pan must be less
than twice the quantity which was indicated on the display before adding
3. current quantity of parts must contain within a tolerance ± 0,3 of the total value,
4. stable measurement result.
If a user claims, that standard quantity is sufficient, it is possible to save the mass of a
single part in balance’s memory by pressing RIGHT ARROW KEY on balance’s overlay.
Fig. 32. Automatic Accuracy Correction saving single part mass in the
database
Set the cursor next to a desired field and insert name for weighed parts. Press Enter key
(for saving the name) and Enter key (for saving the value). Next to the inserted name there

Table of Contents

Other manuals for RADWAG AS series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS series and is the answer not in the manual?

RADWAG AS series Specifications

General IconGeneral
DisplayLCD with backlight
Maximum CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
CalibrationInternal or external, depending on model
Stabilization Time≤ 3 seconds
InterfaceRS232, USB
Repeatability (Standard deviation)Varies by model
LinearityVaries by model
Net WeightVaries by model
Protection ClassIP 43

Related product manuals