RAEGuard 2 PID User’s Guide
36
Bus Baudrate
Choose from three different baud rates for Modbus data transmission:
4800, 9600, or 19200.
1. Press [-] to place the cursor on the baud rate you want.
2. Press [+] to make the selection.
3. Press [MODE] to save your selection.
4. Press [+] to store the selection and exit the submenu and
return to the Monitor Setup menu.
Note: If you do not want to save the change and want to
make further changes, press [MODE]. Press [-] to exit
without saving the changes.
Analogout 4mA
This feature sets the analog output to 4mA for system setup and
troubleshooting.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again
“wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all three digits of the new value are
entered.
3. Press [MODE] to save the 4mA analog output value.
Analogout 20mA
This feature sets the analog output to 20mA for system setup and
troubleshooting.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again
“wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all four digits of the new value are
entered.
3. Press [MODE] to save the 20mA analog output value.