Instrument Assembly for Board and Module Exchange R&S SM300
Version 04 42
Service Manual SM300: 1147.1652.00 Instrument Part No SM300: 1147.1498.03
1.3.3.8 Duoblind
Removing the
duoblind
1. Remove both protectors (see section 1.3.3.1).
2. Remove handle (see section
1.3.3.2).
3. Remove rear frame and both cases (see section
1.3.3.3).
4. Remove carrier lid (see section
1.3.3.5).
5. Undo screws for duoblind, located at bottom of instrument (see
Figure
1-20).
6. Pull duoblind out toward rear of instrument (see
Figure 1-21).
Caution
Module clamp can snag on adjacent module!
Installing the
duoblind
1. Push duoblind into instrument from rear.
Caution
Module clamp can snag on adjacent module
2. Fasten screws located at bottom (DIN7985 M3×4).
3. Install carrier lid (see section
1.3.3.5).
4. Install both cases and rear frame (see section
1.3.3.3).
5. Install handle (see section
1.3.3.2).
6. Install both protectors (see section
1.3.3.1).
Figure 1-20 Figure 1-21
Note
Spare part 1147.1375.00 consist of a monoblind and a duoblind. For the
SM300 only the duoblind is needed. The monoblind can be used for FS300.