R&S SM300  Measurement Uncertainty 
Version 04  71   
Service Manual SM300: 1147.1652.00               Instrument Part No SM300: 1147.1498.03 
2.3  Measurement Uncertainty  
Perform a tolerance 
analysis 
Due to the specifications of the R&S SM300, the measuring equipment must 
meet stringent requirements. Since the required measurement uncertainty of 
the equipment depends on the test setup used, a tolerance analysis needs to 
be performed. 
Tolerances in the 
test report 
To  be  able  to  trace  back  errors  in  the  measurement,  the  measurement 
uncertainty should also be recorded in the test report. 
The tolerances specified in the test report refer to the values specified in the 
data sheet, i.e. the measurement uncertainties of  the test setup used are to 
be taken into account as well. 
Unless otherwise specified, the specified  tolerances indicated are always  to 
be observed. 
  Note 
Please take into account the ETSI Technical Report ETR 028. The specified 
tolerances refer to the R&S SM300 data sheet. 
2.4  Manual Test Procedure 
Measurement 
performance 
The suggested frequencies and levels at which the measurements should be 
performed  have  been  selected  according  to  the  instrument  concept.  Other 
frequencies  and  levels  can  also  be  selected  within  the  scope  of  values 
mentioned in the data sheet. 
  Note 
Before testing the rated specifications, allow the instrument to warm up for at 
least 15 min. The ambient temperature should be 23° C to 26° C. 
Before starting a 
test 
Before starting a test, use the Preset function of the R&S SM300 to place the 
instrument in a defined condition. 
Tested
specifications 
Several  specifications  in  the  data  sheet  are  not  tested.  Due  to  their  digital 
implementation,  it  is  not  necessary  to  test  them  directly.  All  these 
specifications  are  indirectly  taken  into  account  with  the  following 
measurements. 
2.4.1 Self test 
Self test before 
testing performance 
The  R&S  SM300  has  a  self  test  mode  which  checks  that  the  instrument  is 
functioning  properly  and  assists  in  troubleshooting.  Before  the  performance 
test is carried out, the self test should be called up and performed. If the self 
test fails, the following tests are for failure analysis only.