Manual Test Procedure  R&S SM300 
Version 04  76   
Service Manual SM300: 1147.1652.00                Instrument Part No SM300: 1147.1498.03  
Measurement 
1.
Set the levels of two signal generators according to the specifications of 
the used mixer. 
2. For  calibration  purposes  reduce  level  of  EUT  by  40  dB  and  detune  a 
signal generator by 20 kHz. Test signal for harmonics,  the 2nd and 3rd 
harmonic should be more than 30 dB below the fundamental.  Measure 
and note reference value at 20 kHz on analyzer.  
3. Revoke detuning and establish phase quadrature. To do this, set level of 
EUT  again  and  detune  phase  offset  on  auxiliary  generator.  Observe 
mixer output voltage on oscilloscope until the voltage becomes 0. 
4. Read noise voltage on analyzer that  is normalized to a bandwidth of  1 
Hz (noise level). 
Evaluation 
Example 
Form  the  difference  to  the  reference  level  and  add  6  dB  for  the  measured 
(correlated) second sideband and 40 dB to level switching. If the noise level 
of  the  second  signal  generator  is  not  at  least  10  dB  better  than  that  of  the 
EUT,  the  noise  component  of  the  reference  transmitter  too  has  to  be 
determined and calculated.  
The value found gives the correct noise level. 
The reference level is to be measured at 12 dBm. At 20 kHz a noise level of -
78  dBm  (1  Hz)  is  determined.  The  difference  is  90  dB.  In  addition  to  the 
correction for the second sideband  (6  dB)  and  the  level switching (40 dB) a 
noise level of -136 dB or of -136 dBc (dB with reference to the carrier power) 
is obtained. If two identical signal generators are used,  the  result  has  to  be 
reduced  by  3  dB  for  the  (uncorrelated)  noise  power  of  the  reference 
transmitter.
The final result is then -139 dBc.  
2.4.5.4  I/Q Modulation Frequency Response  
Test equipment 
Spectrum analyzer (Table 2-1, item  2) 
Signal generator (
Table 2-1, item 3). 
Test setup 
Connect  the  Rf  output  of  the  SM300  to  the  spectrum  analyzer,
connect the signal generator to the I input of SM300. 
Test method  
By  applying  a  sinewave  AC  voltage  to  the  I  (or  Q)  input,  an  amplitude 
modulation with a suppressed carrier is generated. The modulation frequency 
response  is  determined  by  measuring  the  sidebands  as  a  function  of  the 
frequency of the applied AC voltage.